ISC040-R3 3 ml ready-to-use (RTU)
ISC040-R7 7 ml ready-to-use (RTU)
ISC040-C02 0.2 ml concentrated
ISC040-C05 0.5 ml concentrated
CD68 (Cluster of Differentiation 68) is a transmembrane glycoprotein which is highly glycosylated in its extracellular domain and has a molecular weight of 110 kDa. It is found in the cytoplasmic granules of a number of different blood cells and myocytes and is considered to be a marker for cells of the macrophage lineage, including monocytes, histiocytes, giant cells, Kupffer cells, osteoclasts and their precursors.
The antibody recognizes myelomonocytic and histiocytic tumors and may help distinguish malignant fibrous histiocytomas from other pleomorphic sarcomas.
The stated primary antibody is suitable for immunohistochemical staining of FFPE tissue sections based on specific antigen-antibody reaction. Using a detection system linked to HRP or alkaline phosphatase the antigen visualization is performed via specific binding of the primary antibody. Secondary antibody is binding to the primary antibody, and the enzyme complex labels this complex. The enzymatic activation of the chromogen results in a visible reaction product at the antigen site. Each step is incubated for a precise time and temperature and requires interposed washing steps. The specimen may then be counterstained. Results are interpreted using a light microscope.
The immunostaining procedure causes a colored reaction product to precipitate at the antigen sites localized by the primary antibody.
Cellular localization: Cytoplasm.
A qualified pathologist experienced in immunohistochemistry procedures must evaluate positive and negative tissue controls before interpreting patient specimens.
Positive staining intensity should be assessed within the context of any background staining of the negative reagent control.
Note: A negative result means that the antigen in question was not detected, but not that the antigen is not present in the cells/tissues tested. An antibody panel may be used to support the results in some circumstances. Additionally, the morphology of each tissue sample should be examined utilizing a hematoxylin and eosin stained section. A qualified pathologist must interpret the patient’s morphologic findings and pertinent clinical data.
See MSDS.
| Primary antibody | CD68 |
| Host | Mouse |
| Subclass | IgG |
| Clone | ISM075 |
| Species Cross-reactivity | Human. Others-not known |
| Applications | Immunohistochemistry |
| Epitope Retrieval | Heat-induced epitope retrieval |
| Ready-to-use antibody | Prediluted antibody in antibody diluent buffer |
| Recommended working dilution range | 1:50 to 1:100 |
Product label shows the specific lot number.
Prediluted antibody is ready-to-use and optimized for staining. No further dilution, reconstitution, mixing, or titration is needed.
Antibody concentrate is optimized for dilution within dilution range using InSituChem® Antibody Diluent for IHC (Cat. No. D005-50). Indicated dilution range should be considered as recommendation and depends on different facts (tissue, fixation, incubation conditions, etc.). Optimum dilution to be determined in user’s own system.
The following materials may be required for staining but are not provided with the primary antibody.
– Positive and negative controls
– Microscope slides (positively charged) and cover slips
– Water bath
– Humidified chamber
– Staining jars
– Stopwatch
– Xylene or xylene substitute
– Ethanol
– Deionized or distilled water
– Antigen retrieval reagent, e.g. Antigen Enhancer (HIER buffer), Cat. No. D004-500
– Detection system, e.g. IHC Complete Detection system (Goat anti mouse/rabbit HRP, DAB staining), Cat. No. D001-15
– Wash buffer: e.g. IHC Wash Buffer, Cat. No. D003-500
– Tap water/bluing reagent (e.g. ammonia water)
– Light microscope
Store at 2 – 8 °C.
When stored correctly antibody is stable to the expiration date indicated on the vial. Do not use after expiration date.
To ensure proper reagent delivery and stability of the antibody, replace the dispenser cap after every use and immediately place the bottle cool in an upright position.
[1] Holness CL and Simmons DL (1993): Molecular cloning of CD68, a human macrophage marker related to lysosomal glycoproteins. Blood. 81(6):1607-13.
[2] Warnke RA, Pulford KA, Pallesen G et al. (1989): Diagnosis of myelomonocytic and macrophage neoplasms in routinely processed tissue biopsies with monoclonal antibody KP1. Am J Pathol. 135(6):1089-95.
[3] Pulford KA, Rigney EM, Micklem KJ et al. (1989): P1: a new monoclonal antibody that detects a monocyte/macrophage associated antigen in routinely processed tissue sections. J Clin Pathol. 42(4):414-21.
[4] Gottfried E, Kunz-Schughart LA, Weber A et al. (2008): Expression of CD68 in non-myeloid cell types.
Scand J Immunol. 67(5):453-63.
[5] Athanasou NA, Puddle B, Quinn J and Woods CG (1991): Use of monoclonal antibodies to recognize osteoclasts in routinely processed bone biopsy specimens. J Clin Pathol. 44(8):664-6.
2024-06-10